How to use this document
Copy the template below into your quality system. Replace every <<FILL: ...>> placeholder with your company, system, and process specifics, then route it through your own document control, review, and approval. The filled specimen shows one worked example; adapt the numbers and scope to your operation and verify each cited regulation or standard against the current source before you rely on it.
This is a generic, originally worded starting point. It contains no confidential, proprietary, or employer material, it is drafted with AI assistance and reviewed, it is not legal or compliance advice, and adopting it does not by itself create or prove compliance. See the Disclaimer and Terms.
This is a ready-to-use risk assessment for deciding whether a sterilizing-grade filter’s membrane chemistry is likely to adsorb a meaningful amount of protein, polysorbate, or another formulation component out of the product, and whether a validated pre-flush or a different membrane material is warranted. Replace every <<FILL: ...>> placeholder with your own specifics and route it through your normal document control, review, and approval. A worked filled specimen follows. Verify each cited reference against the current source before you rely on it.
1. Purpose
This assessment evaluates the risk of non-specific adsorption of active ingredient or a critical excipient to the sterilizing filter membrane during final filtration, and documents whether the manufacturer’s stated flux and retention performance alone are sufficient or whether a compatibility study and a validated pre-flush are required.
2. Scope
3. Responsibilities
4. Method and risk factors
Score each factor below as present or absent for the specific formulation and process step. More than one factor present shifts the assessment from “monitor” toward “test,” and any factor present at a protein concentration below roughly 10 mg/mL, or for a surfactant controlled to a narrow specification range, is enough on its own to warrant a small-scale compatibility study before the membrane material is locked in.
5. Small-scale compatibility study summary
6. Conclusion and pre-flush basis
7. References
ASTM F838, Standard Test Method for Determining Bacterial Retention of Membrane Filters Utilized for Liquid Filtration.
PDA Technical Report No. 26, Sterilizing Filtration of Liquids.
BioPhorum Extractables Protocol (current edition), for the related but distinct extractables and leachables assessment.
Confirm the current version of each reference before issue.
8. Revision history
9. Approvals
Filled specimen
Extract for an example monoclonal antibody drug product formulated at 5 mg/mL with polysorbate 80 at 0.02% w/v, filtered through a 0.2 micron hydrophilic PVDF sterilizing filter before fill. Company and product are illustrative.
Small-scale study: filtrate sampled at 5, 15, 30, and 60 L/m2 throughput, compared to a steady-state sample at 150 L/m2. Polysorbate concentration in the first 5 L/m2 sample measured 78% of target, recovering to within specification (95 to 105% of target) by 20 L/m2. Protein concentration and potency showed no significant deviation from target at any sampling point. Conclusion: moderate adsorption risk specific to polysorbate, low risk to protein. Validated pre-flush volume set at 30 L/m2 with margin above the 20 L/m2 recovery point, run with a placebo formulation matching the product’s buffer and polysorbate composition without active ingredient. Batch record specifies the 30 L/m2 pre-flush, discarded to waste, before in-process product sampling begins at the filter outlet.
Common inspection findings this risk assessment prevents
- A filter selected on flux and pressure-drop performance alone, with no documented check of adsorption behavior against the actual formulation.
- A pre-flush performed inconsistently or by informal practice, with no validated volume or batch record language specifying it.
- Polysorbate loss discovered only as an out-of-specification result on a finished batch, with no upstream compatibility data to explain it.
- A low-concentration presentation of an existing product filtered through the same membrane validated at a higher concentration, with no reassessment of adsorption risk at the new concentration.
How to adapt this risk assessment
- Set your document number, product, and membrane identifiers in the header.
- Score every factor in section 4 honestly; a single “Yes” at low concentration or a tightly controlled surfactant specification is enough to warrant the study in section 5.
- Size the pre-flush volume in section 6 with real margin above the point where the study shows recovery, not at the exact recovery point.
- Route the validated pre-flush volume and fluid into the actual batch record language, not just this assessment, so manufacturing executes it consistently.
- Confirm every reference in section 7 against the current published version before issue.